A generation of sensors based on luminescence lifetime
Click here to view the current Data Sheet.
Benefits
Applications
The measurement principle of the sensor system is based on the detection of oxygen concentration as a function of luminescence lifetime, either in dissolved or gaseous phase environments.
Conventional fiberoptic oxygen sensor systems based on intensity measurements are limited in their accuracy by light source stability and ambient light fluctuations. Using a luminescence lifetime detection, measurements are not affected by light source stability, intensity fluctuations caused by fiber bending or changes of the optical properties of the sample (turbidity, refractive index, coloration, etc.).
Calibration: The sensors can be calibrated by a simple two point calibration, 100% air-saturation and 0% air saturation.
The OxiMini and OxyMicro fiber optic oxygen meters are compact, easy to transport. Designed for in/outdoor use, they can be connected to a PC via a RS232 interface. Data can be visualized, analyzed and stored with supplied software. A full range of sensors are available.
The OxyMini is a one channel fiber optic O2 meter for fiber optic O2 minisensors. These sensors are based on 2 mm polymer optical fibers and have a length of 2.5 m.
MiniTip – This dipping probe (501641) has a tip diameter of 4mm and consists of a polymer optical fiber with an O2 sensitive coating. Its range is 0–100%. It has a response time (t90) of ~ 40 s.
MiniFlow – The MiniFlow probe (501642) is a miniaturized fiber optic chemical sensor integrated in a standard T-shape flow through cell which can be easily connected via Luer-Lock adapters to external tubings. Liquids can be pumped through the cell. It has a response time (t90) of approximately 40 s and an excellent long-term stability.
MiniFoil – Sensor material on a 1cm2 support disk made of polyester can be glued inside glass vials. Measure oxygen concentration non-invasively, non-destructively from outside through the wall of the vial. Illuminate the sensor foil with a plastic fiber optic cable (501644, 01645). The wall of the flask must be transparent/non-fluorescent. (Response time (t90) of ~50s.) The material can be implanted into animal tissues or custom-made housings.
The OxyMini system is optimized for process control and biotechnology applications. The measurement principle of the sensor system is based on the detection of oxygen concentration as a function of luminescence lifetime either in dissolved or gaseous phase environments. The fiber optic oxygen meter is compact and easy to transport. It was designed for in/outdoor use and can be connected to a PC via a RS232 interface. Data can be visualized, analyzed and stored with the supplied software. A full range of sensors covering most biomedical applications are available.
Measurement Principle
Conventional fiber-optic oxygen sensor systems based on intensity measurements are limited in their accuracy by light source stability and ambient light fluctuations. Using a luminescence lifetime detection, measurements are not affected by light source stability, intensity fluctuations caused by fiber bending or changes of the optical properties of the sample (turbidity, refractive index, coloration, etc.). These advantages make WPI"s OxyMini and OxyMicro the most advanced and reliable fiber-optic oxygen system available.
Calibration
The sensors can be calibrated by a simple two point calibration, 100% air-saturation and 0% air saturation.
OxyMini Instruction Manual
MiniTip | MiniFlow | MiniSpot | MicroTip | MicroFlow | MicroImplant | |
MEASURE RANGEDISSOLVED/GASEOUS | 0-45 ppm, 0-100%0-760 mmHg | 0-45 ppm, 0-100%0-760 mmHg | 0-45 ppm, 0-100%0-760 mmHg | 0-45 ppm, 0-100%0-760 mmHg | 0-45 ppm, 0-100%0-760 mmHg | 0-45 ppm, 0-100%0-760 mmHg |
RESPONSE TIME [T₉₀]DISSOLVED/GASEOUS | 40 s10 s | 40 s10 s | 40 s10 s | < 2 s< 0.5 s | < 2 s< 0.5 s | < 2 s< 0.5 s |
STERILIZATIONETOH, H₂O₂AUTOCLAVABLE* | YN | YY | YY | YN | YY | YY |
DRIFT** | < 0.1% | < 0.1% | < 0.1% | < 0.3% | < 0.3% | < 0.3% |
ACCURACY*** | 0.2% | |||||
RESOLUTION*** | 2.75 ±0.01 ppm, 9.00 ±0.05 ppm, 220 ±0.15 ppm, 45.0 ±0.25 mmHg,150 ±0.75 mmHg, 375 ±2.6 mmHg | |||||
TEMP RANGE | -10°C to 50°C | |||||
PROBE ASSY LENGTH | 2.5 m | |||||
*130°C, 1.5 atm **100,000 data points, 20°C ***20°C |
WPI(World Precision Instruments Inc.)公司的使命是给神经生理学、心血管生理学和细胞生物学等广大 生物医学科研工作者提供研究用的仪器和工具,使其研究更完善,更**,更精确。WPI China将秉承WPI的理念为中国大陆,中国香港和台湾地区科学 研究者提供优质的产品和周到的服务。
近年来,WPI产品线由神经电生理产品逐渐延伸到生物感应器,心血管生理、病理和心血管药理,光谱检测等领域,革新的产品包括:
WPI-TBR1025/4100, 一氧化氮、硫化氢和过氧化氢等自由基检测仪为全球**和行业领导供应商。
EVOM2, 一款可以在6孔、12孔、24孔和96孔培养板中进行跨膜电阻测量的跨膜电阻仪,是该领域的产品风向标,用于上皮、内皮和肿瘤细胞培养过程中细胞状态的检测,并常用于药物动力学研究和肿瘤药效检测。
WPI-LWCC, 世界上**可以将光程做到2米以上微量样本检测流通池的生产厂家,可以用于海洋微量元素和环境微量物质的检测。
微量注射系统:PV系列气动微量注射系统用于质粒转染和细胞内注射;纳升微量注射系统可以确保您注射的精确到纳升级;多款微注射系统任您选择。
WPI的历史超过40年,在美国、英国、德国、加拿大、澳大利亚、新西兰、法国等欧美生物医学实验室是****的领导供应商之一,供应实验室产品种类超过5000种。我们也将为国内的科学家提供包含神经电生理、生物感应器、心血管和光谱等成套设备的销售及服务,同时提供数据采集系统 、动物实验等手术器械和玻璃毛细管等耗材的销售及服务。